Electronics meets the challenges of patient monitors

Tamara Schmitz, Intersil Corp.

8/22/2010 10:22 PM EDT

The medical industry is keeping pace with other major markets in terms of integration. Single units such as blood pressure cuffs, ECGs and oxygen sensors have successfully entered consumer markets, but there is increasing interest in patient monitors which encompass all of those monitoring abilities and more.

The medical market is generally grouped into three major subcategories: home, clinical and imaging. Home medical is predominately lower cost, portable systems that have lower performance requirements. Clinical and hospital grade equipment features generally higher performance and is more expensive. Imaging systems are generally very large mainframe bases systems, with the exception of ultrasound, that is rapidly expanding into portable and cart-based systems.

Most of these systems are analog-sensor measurement systems, but are applied to biometric functions such as blood pressure, body temperature or heart rate. These biometric sensors are designed to measure physical events such as temperature, pressure, light and flow. After measurement, the system converts them into a corresponding voltage or current. The signal is then conditioned and digitized for processing and analysis. On the control side, the signal is converted back to a current or voltage and applied to an actuator to control such things as air flow, oxygen or temperature.

A typical high-end patient monitor system has five basic subsystems: ECG; pulse oximetry; blood pressure: body temperature; and respiration, Figure 1. Typically, the most critical components in each system are the sensor circuits.


Figure 1: Block diagram of a patient monitor.
(Click on image to enlarge)

Each module uses a different sensor and signal-conditioning circuit. For example, the ECG uses electrodes to measure the electric pulse from the heart. The pulse oximetry (SpO2) uses a light-emitting diode and light sensor to measure oxygen content. Blood pressure is typically measured using a piezo-resistive pressure transducer. For simplicity, several of these biometric modules may utilize common digital, power, and IO subsystems.

Blood Pressure
In blood-pressure biometric modules, the most critical function is the pressure-sensor circuit. Here, precision amplifiers are used to detect very small signals from the transducer and amplify them to a level suitable for ADC processing.

This is typically followed by an active filter to limited unwanted noise at higher frequencies. Amplifiers with low noise, low drift and high gain are necessary to minimize measurement errors and ensure accurate readings. See Figure 2 for a system block diagram.


Figure 2: Block diagram of a blood-pressure system.
(Click on image to enlarge)

The most commonly used piezo-resistive silicon-based pressure sensor in medical applications is the Wheatstone bridge. The pressure-sensing element combines resistors and an etched-diaphragm structure to provide an electrical signal which changes with pressure. As the diaphragm moves under pressure, stress is concentrated in specific areas of the silicon element.

The result is a small voltage that changes in proportion to the pressure applied to the diaphragm. This bridge signal is then amplified using precision op amps prior to ADC conversion.

Key questions to ask when recommending an amplifier are: what is the required accuracy and what are the required voltages? Hospital-grade equipment has different requirements than portable home-based systems, while pressure sensors have varying sensitivities and voltage requirements. The amplifier will generally be selected to match the requirement of the sensor. The ISL28127 and ISL28217 are excellent op amps for ±5V pressure sensor amp gain front-ends due to their low noise and low DC offset and drift.

Electrocardiogram (ECG)
There are several precision amplifier and instrumentation amplifier opportunities in ECG applications. Figure 3 shows a diagnostic (or clinical) ECG with up to 12 leads. Key blocks for lead devices are the electrode gain amplifier, high-pass filter (usually 0.5 Hz), low-pass filter (around 150 Hz) and right-leg drive circuit. For ECG, each electrode requires a precision instrumentation amp to extract a very small signal that rides on a 300 mV to 700 mV common-mode voltage.

Typically, this amplifier will use a higher supply voltage to enable a high gain without railing the amplifier in the presence of the high common-mode voltage from the body. This amp can be a discrete instrumentation amplifier or an integrated instrumentation amplifier. Second- and third-stage active-filter amplifiers are needed to set a very specific band (0.5 Hz – 150 Hz) to capture the ECG QRS wave signal. Typically these will be low-noise, 5V amplifiers with good appropriate bandwidth. In addition, low-noise, low-power amplifiers are needed for the right-leg-drive feedback function.


Figure 3: Block diagram of an ECG system.
(Click on image to enlarge)

In multi-channel systems, such as a 12-lead ECG monitor, it is common to multiplex signals into a common ADC. The key typical requirements for the multiplexer (mux) are low on-resistance and low charge injection.

Generally a specific mux is selected to match the voltage requirements of the filter amplifiers and the ADC. It is also common for multichannel ECGs to have automated lead detection to enable multi-configuration operations. Generally, a low on-resistance switch is used in this circuit as well.

Multiplexers like the ISL43681 and ISL43640 are excellent choices for medical devices as they can operate from 2 V to 12 V. In addition, they have low on-resistance of 39 to 60 Ω, which lowers distortion and reduces ‘kick-back’ voltage from the ADC. Their low charge injection of 0.3 pC to 2 pC (picocoulombs) reduces error contribution on charge-redistribution ADCs. They are also low power and available in the very small QFN package.

For 3-lead portable ECG applications, Iow-power instrumentation amps from a CMOS process may be a better selection. Since CMOS inputs naturally provide a high-impedance input, the need for external buffers is eliminated, saving cost. A good choice for the input amplifier can deliver active feedback with a very precise base line compensation voltage, low 1/f noise, extremely low offset, and low drift versus temperature.

Pulse Oximetry (SpO2)
Oxygen is carried in the blood in hemoglobin and is one of the key vital signs needing detection. Pulse oximetry takes advantage of the fact the blood absorbs certain wavelengths of light differently when it is oxygenated compared to when it is oxygen-deprived.


Figure 4: Block diagram for pulse oximetry system.
(Click on image to enlarge)

The wavelength that marks the difference in absorption to identify oxygen concentration is approximately 805 nm (nanometers) for adult hemoglobin. Therefore, we will use two other wavelengths–one above and one below–to calculate the percentage of oxygenated red blood cells. Usually, 660 nm and 940 nm are used.

A high-impedance, low-bias-current op amp is needed to process the photodiodes that receive the signal at these wavelengths, Figure 4. The ADC also needs to have the fast throughput of a 16-bit device. The DC and background noise is subtracted out, and the pulsed signals are then scaled. Extensive over-sampling, filtering, and signal processing eliminate noise such as movement artifacts from the small signals, and allow the pulse rate to be measured.

Conclusion
Advanced medical systems now require a wide range of amplifiers, multiplexers, converters, references, interface products and power supply products. The portable instruments need the lowest power and most meticulous battery management beyond the stringent noise and gain requirements. The non-portable instruments need reliable isolation and patient protection features, in addition to their precise measurement capability.

It is difficult to offer the best solution for all situations. Every tradeoff of power consumption, available supplies, required resolution, portability, noise and more will play into the selection of the key components. Understanding the implications of device choice will ensure that the system safely delivers quality for the lifetime of the system.

About the author
Tamara Schmitz grew up in the Midwest, finding her way west with an acceptance letter to Stanford University. After collecting three EE degrees (BS, MS, and PhD), she taught analog circuits and test-development engineering as an assistant professor at San Jose State University. With 8 years of part-time experience in applications engineering, she joined industry full-time at Intersil Corporation as a principal applications engineer.





t.alex

8/22/2010 11:58 PM EDT

Developing medical applications is always interesting as it poses wide level of challenges.

Regarding this "Hospital-grade equipment has different requirements than portable home-based system", Could the author elaborate more on this?
Are these about temperature, operating voltage range, or precision level?

Sign in to Reply



hm

8/23/2010 11:09 AM EDT

Biomedical instrumentation is evolving. However, it needs to much accelerate this evolution. When you get an opportunity to closely observe high cost biomedical systems, you may find old type part/topology and quite poor or not so friendly firmware.
As mentioned in this article, many electronic modules are common to all. Can group of manufacturers provide general purpose computing electronic module for all different biomedical measurements? After that, user has to purchase only required sensor module and associated software? Can this general purpose module be desktop/laptop so that we can exploit its ubiquitous availability, powerful processing speed, available wired/wireless communication interfaces and plethora of application development tools?
Can sensor module be made self contained with wireless communication (Bluetooth, wireless USB or similar) and wireless charging? Application software will automatically identify all sensor modules and perform auto calibration.
Can biomedical standardization body write common minimum requirement specifications and their interface to computing devices? They may add other smart devices like ipad or smart phones as required.
This will significantly reduce overall cost of healthcare as related to biomedical instrumentation. Also wireless sensors (with one computing platform or networked computing platforms in hospital environment) adds lot to comfort of patients and healthcare professionals.

Sign in to Reply



TamaraSchmitz

8/23/2010 3:31 PM EDT

As for t.alex's question, you are right about main areas of concern between portable and hospital grade equipment. Portable tends to run on batteries (needs charging/discharging circuitry), has been optimized for low power (to save those batteries) and therefore has sacrificed some of its precision when compared to the hospital-grade versions.

Sign in to Reply



Dr. Phil

8/26/2010 10:23 PM EDT

Given the explosion of computing power in the last 20 yrs or so we now have a PDA capable of being a 2D ultrasound machine.

Just buy a ~$250 USB U/S probe and you are on your way. Gradiations of precision are about useless anymore given the signals we monitor in medicine. Most all modern chips are up to the task.

Laptops are king in clinical medicine these days, cheap and capable...just ad a sensor and you have plenty of horse power for just about anything except maybe for MRI/CAT/PET scanning and data extrapolation and smoothing.

Biological specimens past the DNA are rather simple things for metering in today's world. A lot of the devices are frankly over priced due to law suits not tech innovations.

Sign in to Reply



eewiz

8/30/2010 2:55 AM EDT

I think a real growth in medical electronics is yet to happen. It occurs when everyone's health is continually monitored using wearable body area sensor network & Cellphones. The problem is the lack of integration and high power consumption of discrete components because of which a huge battery is required for reasonable lifetime. I wonder why Intersil or other manufacturers dont consider integrating all the components like Multiplexer & Amplifier into a single chip with a low power wireless.

Sign in to Reply



Please sign in to post comment

Navigate to related information

Datasheets.com Parts Search

185 million searchable parts
(please enter a part number or hit search to begin)

Feedback Form